 
 
 
 
 
   
| Measurement parameter | Stability requirement | 
|---|---|
| Intensity variation  |  0.1 % of normalized  | 
| Position and angle accuracy |  1 % of beam  and  | 
| Energy resolution  |  0.01 % | 
| Timing jitter |  10 % of critical  scale | 
| Data acquisition rate |  10  -10  Hz | 
| Stability period | 10  -10  sec | 
 by centroid motion resulting in
 by centroid motion resulting in 
 projected to the vertical phase space at an aperture located at a certain distance from the source point.
The effect of beam instability on flux transmitted through a phase space aperture depends on the time scale of the fluctuation relative to the detector sampling and data integration times. For fluctuation frequencies much larger than sampling and integration rates, the beam distribution is effectively "smeared out" in phase space, increasing its area but not introducing any new noise. The effective beam emittance is thus given by
 projected to the vertical phase space at an aperture located at a certain distance from the source point.
The effect of beam instability on flux transmitted through a phase space aperture depends on the time scale of the fluctuation relative to the detector sampling and data integration times. For fluctuation frequencies much larger than sampling and integration rates, the beam distribution is effectively "smeared out" in phase space, increasing its area but not introducing any new noise. The effective beam emittance is thus given by 
 . Centroid motion of
. Centroid motion of  30 % of the beam size
30 % of the beam size  and divergence
 and divergence  causes only a 10 % increase in
 causes only a 10 % increase in  
 ignoring possible aliasing effects. Fluctuation frequencies in the range of or less than data integration rates are more harmful. In this case, the beam can move relative to the aperture on a sample-by-sample or scan-by-scan basis, introducing new measurement noise and
 ignoring possible aliasing effects. Fluctuation frequencies in the range of or less than data integration rates are more harmful. In this case, the beam can move relative to the aperture on a sample-by-sample or scan-by-scan basis, introducing new measurement noise and 
 is represented by the envelope of emittance ellipse displacements
 is represented by the envelope of emittance ellipse displacements 
 [1] as shown in Figure 1. Centroid motion of
 [1] as shown in Figure 1. Centroid motion of  5 % causes a 10 % increase in
5 % causes a 10 % increase in 
 . Beam motion occurring over periods much longer than measurement times may have no effect on data quality since the beam is essentially stable. This is especially true if the experiment can be realigned or recalibrated between measurements. The most demanding beam stability requirements arise for a fluctuation frequency interval approximately bounded at the high end by data sampling rates and at the low end by data integration and sample scan times, so that beam noise is not averaged out. Noise spikes or infrequent jumps that do not contribute significantly to the RMS noise floor can be harmful for experiments, particularly those employing difference measurements.
Since most 3rd generation light sources feature low beta (
. Beam motion occurring over periods much longer than measurement times may have no effect on data quality since the beam is essentially stable. This is especially true if the experiment can be realigned or recalibrated between measurements. The most demanding beam stability requirements arise for a fluctuation frequency interval approximately bounded at the high end by data sampling rates and at the low end by data integration and sample scan times, so that beam noise is not averaged out. Noise spikes or infrequent jumps that do not contribute significantly to the RMS noise floor can be harmful for experiments, particularly those employing difference measurements.
Since most 3rd generation light sources feature low beta ( 1 m) straights in order to allow for low gap (
1 m) straights in order to allow for low gap ( 10 mm) insertion devices (IDs), and operate at very small emittance coupling (
10 mm) insertion devices (IDs), and operate at very small emittance coupling ( 1 %) values with horizontal design emittances of just a few nm
1 %) values with horizontal design emittances of just a few nm rad, the requirements compiled in Table 1 lead to sub-micron tolerances for the vertical positional and angular stability of the electron beam at the ID source points (
rad, the requirements compiled in Table 1 lead to sub-micron tolerances for the vertical positional and angular stability of the electron beam at the ID source points (
 
  1
1 m,
m, 
 
  1
1 rad) over a large frequency range 10
rad) over a large frequency range 10 -10
-10 Hz.
 Hz.
 
 
 
 
